Recordeu que es poden pujar les aportacions als sexennis de l'ANECA al repositori institucional, sense necessitat d’esperar que s’òbriga la convocatòria. Fer-ho amb antelació pot evitar imprevistos tècnics i facilitar la gestió. Articles: S’han de dipositar mitjançant la passarel·la GREC/RODERIC. Llibres, capítols de llibre o dades de recerca: S’han de dipositar directament a través de RODERIC. ________________________________________ Recordad que se pueden subir las aportaciones a los sexenios de la ANECA al repositorio institucional, sin necesidad de esperar a que se abra la convocatoria. Hacerlo con tiempo puede evitar imprevistos técnicos y facilitar la gestión. Artículos: Deben depositarse mediante la pasarela GREC/RODERIC. Libros, capítulos de libro o datos de investigación: Deben depositarse directamente a través de RODERIC
Journal article Add to Favorites
Handle

Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: pilot scan electron microscope and mechanical studies

Loading...
Thumbnail Image

Publication Date

Reading Date

Journal Title

Journal ISSN

Volume Title

Directors

Abstract

Objectives: Orthodontic mini-implants (OMIs) are increasingly used in orthodontics but can fail for various reasons. This study investigates the effects of OMI design characteristics on the mechanical properties in artificial bone. Material and Methods: Twelve self-drilling OMIs (2 small, 6 medium, 4 large) from 8 manufacturers were tested for their primary stability in simulated medium-high cancellous bone and the risk to fracture in high-density methacrylate blocks. For the assessments of the maximum insertion torque (IT) and torsional fracture (TF) 5 of each OMI were used and for the pull-out strength (POS) 10. The OMIs were inserted with a torque screwdriver (12 sec/360°) until the bottom at 8 mm depth was reached. OMI designs were analyzed with a scan electron mi - croscope (SEM). Results: SEM images revealed a great variation in product refinement. In the whole sample, a cylindrical OMI shape was associated with higher POS (p<0.001) but lower IT (p=0.002) values. The outer and inner OMI di - ameters were design characteristics well correlated with POS, IT and TF values (ranging from 0.601 to 0.961). Greater thread depth was related to greater POS values (r= 0.628), although OMIs with similar POS values may have different IT values. Thread depth and pitch had some impact on POS. TF depended mainly on the OMI inner (r= 0.961) and outer diameters (r=0.892). A thread depth to outer diameter ratio close to 40% increased TF risk. Conclusions: Although at the same insertion depth the OMI outer and inner diameters are the most important fac - tors for primary stability, other OMI design characteristics (cylindrical vs. conical, thread design) may significantly affect primary stability and torsional fracture. This needs to be considered when selecting the appropriate OMI for the desired orthodontic procedures.

Bibliographic reference

Walter, André ; Winsauer, Heinz ; Marcé Nogué, Jordi ; Mojal, Sergi ; Puigdollers Pérez, Andreu. Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: pilot scan electron microscope and mechanical studies. En: Medicina oral, patología oral y cirugía bucal. Ed inglesa, 2013, Vol. 18, No. 5: 804-810

Publisher

Signature

Full item page

Other versions

Export
endnotezotero
Cite
Citations
plumx
0 citations in
Share

Document Viewer

Select a file:
Reload