Exploring hand-portable nano-liquid chromatography for in place water analysis: Determination of trimethylxanthines as a use case
Analytical performance and optimization of figures of merit of a portable nano liquid chromatograph (NanoLC) with UV detection at 255 nm have been established for in place analysis. Methylxanthines: caffeine, theophylline and theobromine were selected as target analytes. A fast lab method based on IT-SPME coupled on linewith capillary liquid chromatograph (CapLC) with diode array detection (DAD)wasemployed for comparative studies. ITSPME and solid phase extractionwere coupled off-line to NanoLC for improving instrumental parameters,mainly detection capacity and selectivity. IT-SPME or SPE/portable NanoLC based methods were superior in terms of chromatographic resolution and organic solvent consumption per sample, around 200 μL vs 10 mL for ITSPME- CapLC-DAD. Limits of detection (LODs) obtained with the SPE/portable NanoLC were 2-10 ng/mL, which can be suitable for testing the presence of the analytes in several environmental waters in the field. As predictable, the lab method provided better LODs, between 0.1 and 0.5 ng/mL. Good linearity was achieved for both methods and precision was similar for them (≤7%). Both systems were tested for the analysis of real water samples with suitable results.