Objectives: This study was undertaken to characterize the mechanical response of bare (as-received) and single-
layer ceramized zirconia abutments with both internal and external connections that have been developed to enhan
-
ced aesthetic restorations.
Material and Methods: Sixteen zirconia implant abutments (ZiReal Post®, Biomet 3i, USA) with internal and ex
-
ternal connections have been analyzed. Half of the specimens were coated with a 0.5mm-thick layer of a low-fusing
fluroapatite ceramic. Mechanical tests were carried out under static (constant cross-head speed of 1mm/min until
fracture) and dynamic (between 100 and 400N at a frequency of 1Hz) loading conditions. The failure location was
identified by electron microscopy. The removal torque of the retaining screws after testing was also evaluated.
Results: The average fracture strength was above 300N for all the abutments, regardless of connection geometry
and coating. In most of the cases (94%), failure occurred by abutment fracture. No significant differences were
observed either in fatigue behavior and removal torque between the different abutment groups.
Conclusions: Mechanical behavior of Zireal zirconia abutments is independent of the type of internal/external
connection and the presence/absence of ceramic coating. This may be clinically valuable in dental rehabilitation to
improve the aesthetic outcome of zirconia-based dental implant systems.
|
|
Jiménez Melendo, Manuel ; Llena Blasco, Oriol ; Bruguera Alvarez, August ; Llena Blasco, Jaime ; Yáñez Vico, Rosa María ; García Calderón, Manuel ; Vaquero Aguilar, Cristina ; Velázquez Cayón, Rocío ; Gutiérrez Pérez, José Luis ; Torres-Lagares, Daniel. Mechanical behavior of single-layer ceramized zirconia abutments for dental implant prosthetic rehabilitation. En: Journal of Clinical and Experimental Dentistry, 2014, Vol. 6, No. 5: 485-490 |