Performance of long modules of silicon microstrip detectors
NAGIOS: RODERIC FUNCIONANDO

Performance of long modules of silicon microstrip detectors

Repositori DSpace/Manakin

Performance of long modules of silicon microstrip detectors

Mostra el registre complet de l'element

Visualització       (517.4Kb)

Exportar a Refworks
    
Barichello, G.; Cervera Villanueva, Anselmo; Gómez Cadenas, Juan José Perfil; Hernando Morata, José Ángel
Aquest document és un/a article, creat/da en: 1998
This note describes the performance of modules assembled with up to 12 silicon microstrip detectors. These modules were built for the instrumented Silicon Target (STAR) that has been installed in the NOMAD spectrometer, Laboratory and test beam results are compared with model predictions. For a module of nine detectors, test beam results indicate a signal-to-noise ratio of 19, a hit finding efficiency of 99.8% and a spatial resolution of 6.0 mu m. Laboratory measurements indicate that modules of twelve detectors exhibit a signal-to-noise ratio of the order of 16.

    Barichello, G. Cervera Villanueva, Anselmo Gómez Cadenas, Juan José Hernando Morata, José Ángel 1998 Performance of long modules of silicon microstrip detectors Nuclear Instruments & Methods in Physics Research Section A-Accelerators Spectrometers Detectors and Associated Equipment 413 1 17 30
http://dx.doi.org/10.1016/S0168-9002(98)00481-1